Ho:YAG laser lithotripsy in non-contact mode: optimization of fiber to stone working distance to improve ablation efficiency
Article dans une revue avec comité de lecture
Date
2019Journal
World Journal of UrologyRésumé
Purpose: To evaluate how variable working distances between the laser fiber and the stone influence ablation volume. Methods: A laser fiber was fixed on a robotic arm perpendicular to an artificial stone. A single laser pulse was triggered at different working distances (0–2.0 mm in 0.2 mm increments) between the distal fiber tip and the stone. To achieve a measurable impact, pulse energy was set to 2 and 3 J, with either short or long pulse duration. Ablation volume was calculated with an optical microscope. Experiments were repeated five times for each setting. Results: Highest ablation volume was observed with a long pulse of 3 J at a working distance of 0.4 mm between the laser fiber and the stone surface (p value < 0.05). At 2 J, the highest ablation volume was noticed with a short pulse in contact mode. However, ablation volume of the latter was not significantly greater than with a long pulse of 2 J at a working distance of 0.4 mm (p value > 0.05). Compared to lithotripsy in contact mode, triggering a single long pulse at 0.4 mm increased ablation volume by 81% (p value = 0.016) at 2 J and by 89% (p value = 0.034) at 3 J. Conclusions: For Ho:YAG laser lithotripsy, ablation volume may be higher in non-contact mode using long pulses, rather than in direct contact to the stone. Findings of the current study support the need of further studies of lithotripsy in non-contact mode.
Fichier(s) constituant cette publication
Cette publication figure dans le(s) laboratoire(s) suivant(s)
Documents liés
Visualiser des documents liés par titre, auteur, créateur et sujet.
-
Article dans une revue avec comité de lectureUZAN, Audrey; CHIRON, Paul; PANTHIER, Frédéric; HADDAD, Mattieu; BERTHE, Laurent; DOIZI, Steeve; TRAXER, Olivier (2021)Objectives: To compare the risk of laser fiber fracture between Ho:YAG laser and Thulium Fiber Laser (TFL) with different laser fiber diameters, laser settings, and fiber bending radii. METHODS: Lengths of 200, 272, and ...
-
Article dans une revue avec comité de lecturePANTHIER, Frédéric; LAPOUGE, Pierre; DOIZI, Steeve; DRAGOS, L.; BERTHE, Laurent; TRAXER, Olivier (Elsevier Masson, 2020)L’étude in vitro de la lithotritie laser (LL) utilise des calculs humains (CH) ou synthétiques (CS). Devant l’augmentation de la pulvérisation laser (Dusting), peu d’échantillons humains sont disponibles. Les CS sont donc ...
-
Article dans une revue avec comité de lecturePANTHIER, Frédéric; GERMAIN, Thibault; GORNY, Cyril; BERTHE, Laurent; DOIZI, Steeve; TRAXER, Olivier (MDPI, 2022-01)Background: Endocorporeal laser lithotripsy (EL) during flexible ureteroscopy (URS-f) often uses “dusting” settings with “painting” technique. The displacement velocity of the laser fiber (LF) at the stone surface remains ...
-
Article dans une revue avec comité de lectureHADDAD, Mattieu; EMILIANI, Esteban; ROUCHAUSSE, Yann; COSTE, Frédéric; BERTHE, Laurent; DOIZI, Steeve; BUTTICÈ, Salvatore; SOMANI, Bhaskar Kumar; TRAXER, Olivier P. (Springer Verlag, 2017)Purpose: Holmium:YAG laser is the most used laser for urolithiasis. Generally, we use metallic scissors to cut the fiber tip to restore its effectiveness. Many cleaving methods have been described to avoid fiber damage and ...
-
Article dans une revue avec comité de lecturePANTHIER, Frédéric; VENTIMIGLIA, Eugenio; BERTHE, Laurent; CHAUSSAIN, Catherine; DAUDON, Michel; DOIZI, Steeve; TRAXER, Olivier (Springer Verlag, 2020)Introduction: Holmium:yttrium–aluminium–garnet (Ho:YAG) is currently the gold standard for lithotripsy for the treatment of all known urinary stone types. Stone composition and volume are major determinants of the lithotripsy. ...