Comparison of various hemodynamic models for applications to CFD in stent arteries
Communication avec acte
Date
2014Abstract
This work assesses three hemodynamic models for the numerical modeling of intra-stent flows. These are the classical Poiseuille model (PM), the simplified pulsatile model (SPM) and the complete pulsatile model (CPM) based on the analysis of Womersley. They are applied to the physiological flow rate of a stented left coronary artery. The CFD package "Ansys Fluent 14.5" is used to compute the main features of the flows. The results show large differences between the steady and unsteady models notably for the wall shear stress and the re-circulation lengths, which are known to promote intra-stent restenosis. The PM is obviously not pertinent to calculate the flows involved in intra-stent restenosis. The CPM and SPM give close results but the latter model is by far less time-demanding and should be preferred.
Files in this item
Related items
Showing items related by title, author, creator and subject.
-
Article dans une revue avec comité de lectureCHABI, Fatiha; NOGUERA, Ricardo; MAUREL, Blandine; CHAMPMARTIN, Stephane; SARRAF, Christophe (icmmb19.org, 2014)A design study of propeller hydrokinetic turbines is explored in the present paper, where the optimized blade geometry is determined by the classical Glauert theory applicable to the design of axial flow turbines (hydrokinetic ...
-
Article dans une revue avec comité de lectureWe evaluate here three hemodynamic models used for the numerical simulation of bare and stented artery flows. We focus on two flow features responsible for intra-stent restenosis: the wall shear stress and the re-circulation ...
-
Article dans une revue avec comité de lectureMAUREL, Blandine; CHAI, Feng; MATON, Mickael; SOBOCINSKI, Jonathan; HERTAULT, Adrien; BLANCHEMAIN, Nicolas; HAULON, Stephan; LERMUSIAUX, Patrick; SARRAF, Christophe; BAKIR, Farid (Elsevier Masson, 2015)Introduction: In stent restenosis (ISR) remains a major public health concern with an increased morbidity, mortality and health-related costs. Drug-eluting stents (DES) have reduced ISR, but are associated with healing-related ...
-
BrevetL'invention concerne un banc test pour stents, caractérisé en ce qu'il comprend une pompe (2) non volumétrique dont l'entrée est raccordée à un réservoir (4) de liquide sanguin ouvert et dont la sortie est reliée à une ...
-
Article dans une revue avec comité de lectureABBASNEZHAD, Navideh; SHIRINBAYAN, Mohammadali; CHABI, Fatiha; CHAMPMARTIN, Stephane; TCHARKHTCHI, Abbas; BAKIR, Farid (MDPI AG, 2021)Drug-eluting stents are desirable platforms for local medicine delivery. However, the incorporation of drugs into polymers can influence the mechanical and physicochemical properties of said matrix, which is a topic that ...