Finite element analysis of the stump-ischial containment socket interaction: a technical note

Article dans une revue avec comité de lecture
Author
FOUGERON, Nolwenn
ccROHAN, Pierre-Yves
1001017 Institut de Biomécanique Humaine Georges Charpak [IBHGC]
ROSE, Jean-Loïc
1043693 Proteor
BONNET, Xavier
ccPILLET, Helene

URI
http://hdl.handle.net/10985/22116
DOI
10.1016/j.medengphy.2022.103829
Date
2022-07
Journal
Medical Engineering & Physics

Abstract

The role of the above-knee socket is to ensure the load transfer via the coupling of residual limb-prosthesis with minimal discomfort and without damaging the soft tissues. Modelling is a potential tool to predict socket fit prior to manufacture. However, state-of-the-art models only include the femur in soft tissues submitted to static loads neglecting the contribution of the hip joint. The hip joint is particularly challenging to model because it requires to compute the forces of muscles inserting on the residual limb. This work proposes a modelling of the hip joint including the estimation of muscular forces using a combined MusculoSKeletal (MSK)/Finite Element (FE) framework. An experimental-numerical approach was conducted on one femoral amputee subject. This allowed to i) model the hip joint and personalize muscles forces, ii) study the impact of the ischial support, and iii) evaluate the interface pressure. A reduction of the gluteus medius force from the MSK modelling was noticed when considering the ischial support. Interface pressure, predicted between 63 to 71 kPa, agreed with experimental literature data. The contribution of the hip joint is a key element of the modelling approach for the prediction of the socket interface pressure with the residual limb soft tissues.

Files in this item

Name:
IBHGC_MEP_2022_Fougeron.pdf
Size:
1.053Mb
Format:
PDF
Description:
Article principal
View/Open

Collections

  • Institut de Biomécanique Humaine Georges Charpak (IBHGC)

Related items

Showing items related by title, author, creator and subject.

  • Effect of the ischial support on muscle force estimation during transfemoral walking 
    Article dans une revue avec comité de lecture
    ccFOUGERON, Nolwenn; ccBONNET, Xavier; ccPANHELLEUX, Brieuc; ROSE, Jean-Loïc; ccROHAN, Pierre-Yves; PILLET, Hélène; ccPILLET, Helene (Ovid Technologies (Wolters Kluwer Health), 2024-04)
    Background: Transmission of loads between the prosthetic socket and the residual limb is critical for the comfort and walking ability of people with transfemoral amputation. This transmission is mainly determined ...
  • In vivo identification of the mechanical properties of thigh tissues from FreeHand Ultrasound for the numerical investigation of loads at the socket/residual limb interface of amputee people 
    Communication sans acte
    FOUGERON, Nolwenn; BONNET, Xavier; ROSE, Jean-Loïc; ccROHAN, Pierre-Yves; ccPILLET, Helene (2019)
    Prosthetic sockets are custom-designed and are decisive for functionality and comfort of limb prosthesis. To ensure load transmission and stability, high interface stresses are applied. Several computer models of the ...
  • Combining Freehand Ultrasound-Based Indentation and Inverse Finite Element Modelling for the Identification of Hyperelastic Material Properties of Thigh Soft Tissues 
    Article dans une revue avec comité de lecture
    FOUGERON, Nolwenn; HEARING, Diane; ROSE, Jean-Loïc; BONNET, Xavier; ccROHAN, Pierre-Yves; ccPILLET, Helene (American Society of Mechanical Engineers, 2020)
    Finite Element Analysis (FEA) is a numerical modelling tool vastly employed in research facilities to analyse and predict load transmission between the human body and a medical device, such as a prosthesis or an exoskeleton. ...
  • Prediction of muscle forces in residual limb during walking: comparison of transfemoral and Gritti–Stokes amputations 
    Communication avec acte
    FOUGERON, Nolwenn; BONNET, Xavier; PANHELLEUX, Brieuc; ROSE, Jean-Loïc; ccROHAN, Pierre-Yves; ccPILLET, Helene (Taylor and Francis, 2020)
    Evaluation of muscle forces is relevant to understand walking strategies of amputated subjects. Such results could be implemented in finite element modelling to study the interaction between the residual limb and the socket ...
  • Femoral residuum/socket kinematics using fusion between 3D motion capture and stereo radiography 
    Communication avec acte
    PANHELLEUX, Brieuc; FOUGERON, Nolwenn; RUYSSEN, Nicolas; BONNET, Xavier; ccROHAN, Pierre-Yves; ccPILLET, Helene (Taylor and Francis, 2020)
    Measurement of femur motion relative to the socket in gait in TF amputated patients can be a good indicator of prosthesis outcome. Our model, using low dose bi- planar radiography and motion capture, gives a prediction of ...

Browse

All SAMCommunities & CollectionsAuthorsIssue DateCenter / InstitutionThis CollectionAuthorsIssue DateCenter / Institution

Newsletter

Latest newsletterPrevious newsletters

Statistics

Most Popular ItemsStatistics by CountryMost Popular Authors

ÉCOLE NATIONALE SUPERIEURE D'ARTS ET METIERS

  • Contact
  • Mentions légales

ÉCOLE NATIONALE SUPERIEURE D'ARTS ET METIERS

  • Contact
  • Mentions légales