The effect of polystyrene sodium sulfonate grafting on polyethylene terephthalate artificial ligaments on in vitro mineralisation and in vivo bone tissue integration
Article dans une revue avec comité de lecture
Date
2013Journal
BiomaterialsAbstract
This study investigates the impact of polystyrene sodium sulfonate (PolyNaSS) grafting onto the osseointegration of a polyethylene terephthalate artificial ligament (Ligament Advanced Reinforcement System, LARS ) used for Anterior Cruciate Ligament (ACL). The performance of grafted and non-grafted ligaments was assessed in vitro by culturing human osteoblasts under osteogenic induction and this demonstrated that the surface modification was capable of up-regulating the secretion of ALP and induced higher level of mineralisation as measured 6 weeks post-seeding by Micro-Computed Tomography. Grafted and non-grafted LARS were subsequently implanted in an ovine model for ACL reconstruction and the ligament-to-bone interface was evaluated by histology and biomechanical testings 3 and 12 months post-implantation. The grafted ligaments exhibited more frequent direct ligament-tobone contact and bone formation in the core of the ligament at the later time point than the nongrafted specimens, the grafting also significantly reduced thefibrous encapsulation of the ligament 12 months post-implantation. However, this improved osseo-integration was not translated into a significant increase in the biomechanical pull-out loads. These results provide evidences that PolyNaSS grafting improved the osseo-integration of the artificial ligament within the bone tunnels. This might positively influence the outcome of the surgical reconstructions, as higher ligament stability is believed to limit micro-movement and therefore permits earlier and enhanced healing.
Files in this item
Related items
Showing items related by title, author, creator and subject.
-
Article dans une revue avec comité de lectureVIATEAU, Véronique; MANASSERO, Mathieu; ANAGNOSTOU, Fani; MITTON, David; MIGONNEY, Véronique; GUÉRARD, Sandra (WB Saunders, 2013)PurposeThe purposes of this study were to assess tissue ingrowth within the Ligament Advanced Reinforcement System (LARS) artificial ligament (LARS AC; LARS, Arc sur Tille, France) and to study the biomechanical characteristics ...
-
Article dans une revue avec comité de lectureGUÉRARD, Sandra; MANASSERO, Mathieu; VIATEAU, Véronique; MIGONNEY, Véronique; SKALLI, Wafa; MITTON, David (techno-press, 2014)This study aimed to assess the biomechanical performance of a new generation of artificial ligament, which can be considered “bioactive” and “biointegrated,” implanted in sheep. Thirty sheep were implanted: 15 sheep received ...
-
Article dans une revue avec comité de lectureMAEZTU REDIN, Deyo; CAROUX, Julien; ROHAN, Pierre-Yves; CERMOLACCE, Alexia; TRNKA, Julien; MANASSERO, Mathieu; VIATEAU, Véronique; CORTÉ, Laurent; PILLET, Helene (Elsevier BV, 2022-09)Impingement with surrounding tissues is a major cause of failure of anterior cruciate ligament reconstruction. However, the complexity of the knee kinematics and anatomical variations make it difficult to predict the ...
-
Communication sans acteSKALLI, Wafa; TRNKA, Julien; MANASSERO, Mathieu; VIATEAU, Véronique; CAROUX, Julien; CORTÉ, Laurent; ROHAN, Pierre-Yves; PILLET, Helene (2017)Introduction Rupture of the Anterior Cruciate Ligament (ACL) affects about 1 person over 3000 every year. The current standard care is based on ligament reconstruction by autograft from tendon tissues and is considered as ...
-
Article dans une revue avec comité de lectureBENOIT, Aurélie; GILLET, Brigitte; GUILLOT, Geneviève; HILD, François; MITTON, David; PERIE, Jean Noël; ROUX, Stéphane; GUÉRARD, Sandra (Elsevier, 2009)A mini-compression jig was built to perform in situ tests on bovine trabecular bone monitored by micro-MRI. The MRI antenna provided an isotropic resolution of 78 microm that allows for a volume correlation method to be ...